$0.00

Temperature Calibrator – 724

In stock
Compare

Description

The new Fluke 724: everything you need to troubleshoot and calibrate temperature.
Now you can carry one tool to expertly test all the temperature sensors and transmitters in your plant.
The Fluke 724 is the master’s answer to temperature calibration.
The Fluke 724 can measure and source 10 thermocouple types and 7 RTD types, plus volts and ohms.
The Fluke 724 even handles high-speed pulsed RTD circuits and provides loop power.
The dual line measure/source display lets you source temperature and view loop current at the same time.
With its simple, “no menus” controls, the Fluke 724 is easy to operate, too.

Product code: Fluke 724

OPTIONS AVAILABLE

Fuke 724, Fluke 724 with IANZ Cal Cer

Technical Information

SPECIFICATIONS
Ramp Functions Source Functions: Voltage, current, resistance, frequency, temperature
Ramps: Slow ramp, Fast ramp, 25% step-ramp
Loop Power Function Voltage: 24V
Accuracy: 10%
Maximum current: 22 mA, short circuit protected
Step Functions Source Functions: Voltage, current, resistance, frequency, temperature
Steps: 25% of range, 100% of range
MEASUREMENT ACCURACY
Voltage DC 30.000 V 0.02% + 2 counts (upper display)
20.000 V 0.02% + 2 counts (lower display)
100.00 mV 0.02% + 2 counts
-10.00 mV to 75.00 mV 0.025% + 1 counts (vis TC Connector)
Current DC 24.000 mA 0.02% = 2 counts
Resistance 0.0 to 400.0Ω (4-wire), 0.15Ω (2- and 3- wire)
401 to 1500Ω  0.5Ω (2- and 3-wire)
1500 to 3200Ω 1Ω 94-wire), 1.5Ω (2- and 3-wire)
SOURCE ACCURACY
Voltage DC 100.00 mV 0.02% + 2 counts
10.00 V 0.02% + 2 counts
-10.00 mV to 75 mV 0.025% + 1 count (via TC Connector)
Resistance 15.0 to 400.0Ω 0.15Ω (exc. current 0.15 to 0.5 mA), 0.1Ω (exc. current 0.5 to 2 mA)
401 to 1500Ω 0.5Ω (excitation current 0.05 to 0.8 mA)
1500 to 3200Ω 1Ω (excitation current 0.05 to 0.4 mA)
FOR FULL SPECIFICATIONS REFER TO MANUFACTURERS DATA SHEET