$0.00

IANZ Certification of Beamex Process Calibrators

In stock
ADT223A-1 ,

Description

Before any calibration is undertaken on your Beamex, we leave the unit in our lab ( at 20 deg C) for 3 hours to come up to optimum operating temperature, then we check your unit to see if its firmware needs upgrading– if it does, we undertake this for you. Also, if the date and time need correcting, we set this for you also.

All of our calibrations are done to so to meet the manufacturer’s specifications for your unit. We will adjust if it falls outside its specified tolerance. If needed, we can take results before adjustment and record these for you ( just ask). Normally, we just record the adjustment amount we made, ie A span adjustment of -0.150 kPa was made to this item.

When CPS carries out the calibration of your Beamex, this is a total solution calibration – we do it all, there is no function we don’t calibrate. When you send it to us we take care of the whole job, and we do this completely within CPS. So you can have faith that the calibration being done is of the highest standard and all calibrations are fully IANZ endorsed. Plus our aim is to do it in 7 days!.

As per all our calibrations, we do a full uncertainty analysis at each calibration point, so you know your inaccuracy tolerance at each and every calibration point.

Sample CPS calibration certificates can be downloaded from this page above.

For the calibration of Beamex Multi Function process calibrators, we use nearly all our reference equipment. For the electrical source and measure functions, we calibrate these using our reference Fluke 8508A multimeter combined with a Fluke 9100E electrical Calibrator. Click on the Technical info tab above to see exactly what we calibrate on the electrical side.

We calibrate all the electrical functions on your Beamex.

For the pressure modules what we use one of our three Fluke pressure controllers. The 400 mbar module is very common, and we do this using one of our PPC4, which is set up for low-pressure calibration work. This has an accuracy of 0.008% of RDG. These pressure controllers are exactly the type of equipment Beamex uses for their final calibration.

At this stage, we can calibrate the internal and external pressure modules of MC5’s automatically over the space of about 3 hours. This is where we use a pressure controller to apply a highly accurate known pressure, and then as this pressure is being applied, we take a number of readings over 60 seconds from the MC5 and average them both to get applied readings and the MC5 readings – we are the only company doing this. For the MC2’s 4’s and the MC6’s, we still use the same software and pressure controllers, but we just take the data manually. Our intention is for us to undertake calibration of all the Beamex units automatically in the near future.

Upon the completion of the full calibrations of your unit, we then go and set the next calibration date internally for you.

 

Technical Information

You may also need